X
Call Us Today! 022 655 443 30 | customercare@genotronics.in

*Genotronics is committed for customer satisfaction and quality. our organization is engaged in Trading, Importing , supplying and service provider we are offering a wide range of Medical Equipment. These equipment are manufactured at hi-tech manufacturing unit under the observations of Expert professionals. Our range is durable, efficient, accurate in results and portable as well. Further these Medical Equipment can be availed from us at reasonable prices.

Trilogy 202

Simplicity and versatility for sub-acute ventilation

Support many needs with a single ventilator

The Trilogy 202 is both a volume-control and pressure-control ventilator for invasive and noninvasive ventilation. The versatile breath delivery and setup options free you from burdensome equipment exchanges, providing greater continuity of care for your patients. Because the Trilogy 202 has the unique ability to compensate for leaks in both pressure and volume control modes, using simpler passive circuits may support significant time and cost savings. With one simple setting change, the Trilogy 202 supports either active or passive exhalation breathing circuits to accommodate changes in circuit preference.

Intra-hospital transport

The Trilogy 202 allows freedom of movement. With a 3-hour internal battery and optional 3-hour detachable battery, you’ll have sufficient power for intra-hospital transports – with plenty of time for any delays or detours. It weighs only 5.6 kg and has an integrated carrying handle for safe and convenient handling during intra-hospital transports.

Hospital to home transition

In addition to sharing a common user interface, all home and hospital Trilogy ventilators offer identical ventilation modes and breathing circuit configurations.

Now your home-bound patients will receive the same finely tuned ventilatory support they received in the hospital. Furthermore, stored data from the home or the hospital can be downloaded for review and evaluated for meaningful trends.

Synchrony

The Auto-Trak algorithm improves patient-ventilator synchrony by adjusting to changing breathing patterns and dynamic leaks. The auto-adaptive triggering, cycling, and leak adjustments may help reduce the time clinicians spend adjusting thresholds and re-fitting masks.

Leak compensated volume modes

Trilogy 202 offers all the conventional volume-control modes: AC, SIMV (with or without PS), and CV. The Trilogy 202 has the unique ability to compensate for leaks in a volume control mode. This means more consistent tidal volume delivery even during NIV or when cuff leaks are present.

 

Spcification

Controls
CPAP
4 - 20 cmH20 (passive leak port circuits) cm H₂O
IPAP
50 cm H₂O
EPAP/PEEP (a)
0 - 25 cmH20 (active valve circuits) cm H₂O
EPAP/PEEP (b)
4 - 25 cmH20 (passive leak port circuits) cm H₂O
Pressure support
0 – 30 cmH2O cm H₂O
Tidal volume
50 - 2000 ml
Breath rate(a)
0 - 60 BPM (AC mode) beats per minute
Breath rate(b)
1 - 60 BPM (all other modes) beats per minute
Inspiratory time
0.3 - 5.0 s Synchrony features
Rise time
1 - 6 (relative scale)
Ramp start pressure (a)
0 - 25 cmH20 (active circuits) cm H₂O
Ramp start pressure (b)
4 - 25 cmH20 (passive circuits) cm H₂O
Ramp start pressure (c)
4 - 19 cmH20 (CPAP mode) cm H₂O
Ramp time Off
5 - 45 min
C-Flex Off
1 - 3 (relative scale)
Flow trigger sensitivity
1 - 9 l/min
Flow cycle sensitivity
10 - 90 %
Environmental
Operating temperature
5 – 40 °C
Relative humidity
15 - 95 %
Storage temperature
-20 – 60 °C
Atmospheric pressure
60 - 110 kPa (450 - 825 mmHg)
Patient types
Adult
Yes
Pediatric
≥ 5 kg or greater
Electrical
Input voltage
100 - 240 VAC and 50/60 Hz and 2.1 A
Detachable battery voltage
14.4 VDC
Internal battery life
3 hours under normal conditions hr
Detachable battery life
3 hours under normal conditions hr
external battery connection
12 VDC
Volume modes
Pressure control (PC)
Yes
MPV Mouthpiece Ventilation Mode
Yes
Synchronized intermittent mandatory ventilation (SIMV)
Yes
SIMV with pressure support (SIMV w/PS)
Yes
Control ventilation (CV)
Yes
Alarms
Circuit disconnect Off
10 - 60 sec
Low tidal volume Off
50 - 2000 ml
Apnea Off
10 - 60 s and 4 - 60 BPM
High tidal volume Off
50 - 2000 ml
High minute ventilation Off
1 - 99 l/min Off, 1 - 99 l/min
Low minute ventilation Off
1 - 99 l/min
High respiratory rate Off
4 - 80 beats per minute
Low respiratory rate Off
4 - 80 beats per minute
Pressure modes
Pressure control (PC)
Yes
Pressure control-SIMV (PC-SIMV)
Yes
Spontaneous ventilation (S)
Yes
Spontaneous ventilation with timed back-up (S/T)
Yes
Timed ventilation (T)
Yes
Continuous positive airway pressure (CPAP)
Yes
Average volume assured pressure support (AVAPS)
With passive circuit type, S, S/T, PC, and T modes only
Options
Detachable back-up battery
Offering up to 3 hours additional operating time
Hospital roll stand
Provides convenient accessory basket and humidifier mount
DirectView
1 GB SD card data storage integrated into the ventilator software
Monitored parameters
Tidal volume
0 - 2000 ml
Minute ventilation
0 - 99 l/min
Estimated leak rate
0 - 200 l/min
Respiratory rate
0 - 80 BPM 0 - 80 beats per minute
Peak inspiratory flow
0 - 200 l/min
Peak inspiratory pressure
0 - 99 cm H₂O
Mean airway pressure
0 - 99 cm H₂O
% patient triggered breaths
0 - 100 %
I:E ratio
9.9:1 - 1:9.9
Synchrony features
Auto-Trak sensitivity
Auto-Trak sensitivity Auto-adaptive triggering, cycling, and leak compensation (available in all modes, passive circuit only)
Adjustable flow triggering
1 – 9 l/min (available in all modes and circuit types) l/min
Circuit types
Active exhalation valve with proximal airway pressure (PAP)
Yes
Active exhalation valve with flow sensor
Yes
Passive exhalation port
Yes
Compliance
IEC 60601-1 Medical electrical equipment Part 1: General requirements for safety
Yes
IEC 60601-1-2 General requirements for safety – collateral standard Electromagnetic compatibility – requirements and tests
Yes
IEC 60601-2-12 Medical electrical equipment Part 2-12: Particular requirements for the safety of lung ventilators – Critical Care ventilators
Yes
Oxygen
FiO2
21 - 100 %
02 flush
2 min at 100%
02 input pressure range
276 - 600 kPa (40 - 87 psi)